Departments of Radiology and Biomedical Engineering, University of Iowa, Iowa City, Iowa
Correspondence and requests for reprints should be addressed to Eric A. Hoffman, Ph.D., Department of Radiology, University of Iowa Carver College of Medicine, 200 Hawkins Drive, Iowa City, IA 52240. E-mail: eric-hoffman{at}uiowa.edu
ABSTRACT
With the emergence of multidetector-row computed tomography (CT) it is now possible to image both structure and function via use of a single imaging modality. Breath-hold spiral CT provides detail of the airway and vascular trees along with texture reflective of the state of the lung parenchyma. Use of stable xenon gas wash-in and/or wash-out methods using an axial mode of the CT scanner whereby images are acquired through gating to the respiratory cycle provide detailed images of regional ventilation with isotropic voxel dimensions now on the order of 0.4 mm. Axial scanning during a breath hold and gating to the electrocardiogram during the passage of a sharp bolus injection of iodinated contrast agent provide detailed images of regional pulmonary perfusion. These dynamic CT methods for the study of regional lung function are discussed in the context of other methods that have been used to study heterogeneity of lung function.
Key Words: computed tomography, functional, quantitative, xenon pulmonary blood flow pulmonary perfusion pulmonary ventilation
The primary role of the lung is to deliver oxygen from the air into the venous blood and to extract carbon dioxide from the blood at the same airblood interface. Thus, matching of ventilation and perfusion within various regions of the lung is critical for adequate gas exchange. To understand the etiology of physiologic manifestations of lung pathology, it has been critical to understand the interrelationship between regional lung ventilation and regional lung perfusion and to interrelate these two functional parameters to lung structure. Imaging methods are emerging to provide powerful new tools in the quest for this understanding.
To infer nonuniform lung ventilation, a large amount of early work was based on simple nitrogen wash-out curve (1) imaging of the distribution of inhaled radioactive xenon gas (2), regional strain measured via implanted lung markers used in conjunction with biplane X-ray fluoroscopy (36), as well as regional density changes seen with X-ray computed tomography (CT) imaging of the lung and multiple static inflation steps (7, 8). For the measurement of pulmonary perfusion, radioactive gases (911), radiolabeled microaggregated albumin, and radioactive microsphere injection (12, 13) have been extensively used. Ventilationperfusion relationships in lung compartments, without specific knowledge of where the compartments are or whether they are contiguous, have been studied via the multiple inert gas elimination technique (14). These early studies provided significant insights into lung function and served to establish the notion that ventilation and perfusion are heterogeneous but well matched under normal conditions. A revolution in X-ray CT methods, taking advantage of multidetector-row CT (MDCT), has now provided a means of quantitatively studying regional lung ventilation and perfusion in conjunction with acquisition of highly detailed information regarding lung structure. Simultaneously there has been a demand for objective, quantitative tools with which to assess regional lung structure and function as new methods for intervening in lung disease emerge. Surgical and nonsurgical interventions in late-stage emphysema have increased the demand for methods by which to identify subsets of patients likely to benefit from the interventions, transbronchial biopsies have increasingly required imaging-based methods of guidance, and therapies targeted at interventions early in the pathologic development of a disease require tools more sensitive than the pulmonary function test or the radiologist's subjective visual reading of a radiograph or CT scan to identify subjects needing early intervention and to evaluate the effectiveness of the intervention. As interest increases in delivering drugs via inhalation rather than injection, it becomes critically importance to be able to evaluate the effect of these drugs, such as insulin, on the lungs. Elsewhere in this issue, the reader will find discussions of magnetic resonance imaging, positron emission tomography (PET), and bronchoscopy, and discussions of methods for assessing anatomic details of the lung, such as airway geometry. In this article, we focus on methods for imaging ventilation (V·), perfusion (Q·), and V·/Q· via CT methodologies, and we place these methods into context with other methods that have provided our current understanding of lung structurefunction relationships.
MDCT
Early studies of the lung via CT imaging used a one-of-a-kind scanner developed at the Mayo Clinic (Rochester, MN), known as the dynamic spatial reconstructor (1517). This scanner provided the means of capturing a volumetric image of the lung in as short a time as 1/60th of a second, with increased resolution provided by widening the scan aperture, and typically apertures of 4/60th of a second were used for assessing regional perfusion, which was calculated by the injection of iodinated contrast agent and scanning gated to the cardiac cycle as the contrast agent passed through the lung with the lung held at a fixed volume (18). Regional ventilation was estimated by assessing regional density changes in the lung as the lung was inflated and imaged at each static inflation step (7, 8). Because early CT scanners required up to 2 to 5 s to acquire projection data for the reconstruction of a single slice of the lung, little work was done to assess lung function in the clinical arena via this modality, and the modality gained a reputation of being strictly targeted at imaging anatomic features. Imatron Corporation (South San Francisco, CA) developed a scanner known initially as the Ultrafast or Cine CT, later referred to as the electron beam CT (EBCT) (19), which provided scan apertures of 50 ms and up to eight slices of 7-mm thickness each or thin, single 3-mm-thick sections gathered in 100 ms. The table step took 1 to 2 s and thus images of the full lung could be acquired during a static breath hold on the order of 40 s. The EBCT scanner had a spatial resolution of approximately 10 line pairs per centimeter. Although this modality provided a useful research tool with which to assess both ventilation (20) and perfusion (21), this application of the EBCT remained impractical for use in the clinical arena. At the end of the 1990s a new class of CT scanners emerged, known as MDCT scanners. These scanners had reduced rotation times of 0.5 s and initially acquired four simultaneous thin (1- to 1.5-mm) sections. Using a spiral mode of scanning, highly detailed anatomic images of the lung could be acquired volumetrically with a breath hold of 40 s, but with an increased spatial resolution on the order of 24 line pairs per centimeter. Because scan aperture (rotation time) was down to 0.5 s it was now possible to image axially and gate the scanner to the cardiac cycle during a breath hold such that blood flow measures could be assessed similarly to those acquired earlier via the dynamic spatial reconstructor (18, 22) or later by the EBCT scanner (2325). These MDCT scanners have rapidly evolved to such as point that they can now acquire 64 slices in a single rotation and a rotation is completed in as little as 0.3 s. Volumetric images are now acquired with a breath hold of 5 s or less and axial blood flow assessment is widened such that up to 4 cm of z-axis coverage is achievable. It is clear that z-axis coverage will soon widen as rotation times will continue to drop. A fully equipped MDCT scanner suite is shown in Figure 1 and includes a xenon delivery device (Enhancer 3000; Diversified Diagnostic Products, Houston, TX), a high-powered contrast injector (MEDRAD, Indianola, PA), and a physiologic monitoring system (IntelliVue; Philips Medical Systems, Bothell, WA). The xenon Enhancer 3000 allows for the rebreathing of xenon gas, the scrubbing of CO2, and maintenance of the O2 and Xe concentration constant. The contrast injector provides 950 to 1,200 psi of pressure to deliver 0.5 cm3 of iodinated contrast agent per kilogram in 2 to 3 s, with injection into the superior vena cava. The inset to Figure 1 shows a micro-CT scanner (microCAT II; Siemens Medical Systems, Knoxville, TN), which allows for imaging mice at 18-µm resolution, such that one can visualize alveoli in vivo. This is briefly discussed at the conclusion of this article.
|
Perfusion Measurement Methods
Although multiple methods are currently available to measure pulmonary perfusion quantitatively, they have relative limitations and are not used routinely in clinical applications.
The technique using radiolabeled red blood cells or serum albumin macroaggregates was applied to evaluate regional pulmonary transit time and blood flow by determining the amount of radiation emitted from a specific region of interest (ROI) (26). Also, regional perfusion can be measured by using insoluble radioactive gas with a low bloodgas partition coefficient, such as 133Xe, on the basis of the concept that radioactivity distributes within alveolar in proportion to local perfusion (27).
Regional pulmonary perfusion has been measured on the basis of the concept that the number of microspheres trapped in an ROI is proportion to regional perfusion (28). After microspheres are injected intravenously, fluorescence intensities are measured by extracting fluorescent dye and dye concentration is determined by fluorimetry. This method has relatively poor spatial resolution (because of the need to acquire enough spheres for a meaningful estimate of flow and because of the difficulty of dicing the lung too finely).
Mijailovich and coworkers (29) provided the theoretical background for positron emission tomographic (PET) measurement of pulmonary perfusion from 13N2 concentration during apnea. When nitrogen gas arrives in the pulmonary capillaries, most of the tracer gas diffuses into the alveolar air space during a first pass because of its low solubility in blood and tissue; regional tracer content during apnea is proportional to regional Q·. Brudin and coworkers (30) calculated pulmonary perfusion by measuring ventilation obtained during continuous inhalation of 19Ne and V·/Q· during constant intravenous infusion of 13N in saline solution. This technique has reduced temporal and spatial resolution, compared with magnetic resonance imaging (MRI) and CT. More recently, Venegas and colleagues (3133) used PET to measure perfusion. Furthermore, with the modeling they employ, they have made perfusion estimates that have allowed them to predict blood gases. Others (34) have used PET to measure perfusion with 15O-labeled water and also to measure perfusion via PET with macroaggregates of labeled albumin (effectively microspheres that degrade in the body and can be used in humans). With the implementation of commercial PET CT scanners, it is now possible to begin to use these modalities to cross-validate each other and to begin to design studies that best make use of the unique capabilities offered by each modality. Furthermore, with the more limited availability of PET scanners, it may be of interest to identify how CT may provide the same information as PET. CT clearly has resolution such that it is able to evaluate flow in, for instance, nodules of sizes not visible in a PET image.
Wolfkiel and Rich (25) measured myocardial blood flow by EBCT, using a single-compartment model of indicator transport. The basic concept of their model is that flow is proportional to maximum enhancement. Wu and coworkers, using the dynamic spatial reconstructor, applied this model in pigs to measure pulmonary blood flow and showed a good correlation with microsphere-based flow measurements. The assumption of the model used is that tissue accumulation of indicator must be nearly complete before the onset of indicator wash-out. Wolfkiel and Rich tested this assumption by comparing timedensity curves of the pulmonary outflow tract, lung parenchyma, and aortic outflow tract. According to their article, even if the arrival time on the left side of the circulation was slightly before the peak lung parenchymal enhancement time, estimation of the amount of wash-out was minimal (2.7 ± 5.3%) at the time of peak lung contrast enhancement, suggesting that the assumption of no indicator wash-out is valid for pulmonary blood flow measurements made with central intravenous bolus injections. This model using CT has been applied by other investigators studying both animals and humans (35). Won and coworkers (36) demonstrated the ability to use dynamic CT imaging to extract the signal within a single voxel related to microvascular blood flow, thus demonstrating the ability to estimate perfusion parameters at or near the very interface that provides gas exchange in the lungs. This was done by assessing the timeattenuation curve in the feeding pulmonary artery and in the peripheral parenchymal regions to then extract the transfer function, which expresses the mathematical function serving to convert the arterial curve into the parenchymal curve. It was observed that the transfer function was bimodal; and it was concluded that if the first mode reflected the partial-volume small arteries in a voxel, and the second mode reflected the partial-volume microvascular bed, then one could eliminate the first portion of the transfer function, reconvolve the arterial curve with just the second portion of the transfer function, and thus derive a timeattenuation curve reflecting microvascular perfusion. An example of perfusion measures obtained in a human nonsmoker and in a human smoker is shown in Figure 2 (p. 506) with scanning having been achieved via MDCT (SOMATOM Sensation 16; Siemens Medical Solutions, Forscheim, Germany). Note the increased heterogeneity of perfusion in the smoker, possibly representing disruption of normal perfusion via inflammatory processes.
Use of pulmonary perfusion measures is largely limited to date to the assessment of pulmonary emboli (3741) or the characterization of lung nodules (42). No study to date has used quantitative measures (ml/min/g) for clinical assessment of the lung. Studies assessing pulmonary emboli and lung nodules have relied on slow, prolonged infusions of contrast simply to evaluate enhancement of the local blood pool. This has erroneously been termed "perfusion." It is in fact "perfused blood volume."
MRI has been used to evaluate pulmonary perfusion in animals and humans (43). Perfusion can be evaluated by analyzing enhancement after intravenous injection of contrast material (typically gadolinium). CT has significant potential for measuring highly regionalized pulmonary blood flow on the basis of first-pass indicator dilution principles, using iodinated contrast medium (44). The measurement of blood flow in brain (45) and myocardium (24), using CT, was introduced in the early 1980s. However, the measurement of blood flow has been limited by the long scan time, long interscan delays, and a lack of nondiffusible indicators (46, 47). The indicator must satisfy several criteria in order to allow the application of first-pass indicator dilution theory: (1) contrast indicator should not affect flow through physiologic or volume effects, (2) the indicator must mix with the blood uniformly, and (3) the indicator must remain in the intravascular space. The development of the EBCT and the introduction of nonionic contrast agents improved the ability of CT to measure regional tissue perfusion.
Ventilation Measurement Methods
Quantification of regional pulmonary ventilation has been attempted via various imaging techniques. Scintigraphic imaging (48) can measure regional ventilation as well as perfusion. Radiotracers, such as 133Xe, 13N, and 15O, have been used tomographically for the evaluation of lung function (49). Nitrogen and oxygen have relatively short half-lives, and specialized imaging equipment is required (50). This technique yields gross estimates of ventilation within relatively large tissue volumes.
PET combined with radioactive inert gas tracer was developed to measure regional alveolar ventilation. Regional ventilation was obtained during the continuous inhalation of 19Ne (51). Senda and coworkers (52) introduced a method based on the inhalation of nitrogen gas instead of 19Ne. PET imaging during a constant rate intravenous infusion of nitrogen has been used to measure ventilationperfusion ratios. Musch and coworkers (31) modified this technique to measure ventilation and perfusion separately by use of a bolus infusion of nitrogen during a short apneic period. This method has several advantages to assess both ventilation and perfusion in topological distribution: (1) use of a single administration of tracer, (2) minimal invasiveness, and (3) low radiation exposure to the subject. However, there are several shortcomings. First and foremost has been the fact that function is not matched with structural detail. This is being solved by the emergence of combined PET-CT scanners whereby the best of both PET and CT can be obtained if one uses state-of-the-art MDCT in conjunction with the PET side of the scanner. Often, however, the CT side is a lower level machine that is there largely to provide attenuation correction information obtained during a fairly prolonged breath hold.
Fluorescent microspheres (12) can be used to measure regional ventilation by inhaling fluorescent microsphere aerosols and measuring regional extracted dyes for each lung portion. The most significant limitation of aerosol deposition as an estimate of regional ventilation is related to the scale of measurement (53, 54). When a 1.9-cm3 volume per piece is used, ventilation relies primarily on convective gas movement, whereas the deposition of 1.0-µm aerosol depends on gravitational settling and inertial impaction, neither of which is related to alveolar ventilation.
MRI of regional lung ventilation has been developed. T1- or T2-weighted proton MRI is significantly hampered by the low concentration of spin within the lung, susceptible artifact, and cardiacrespiratory motion (55). New methods have been introduced to increase the signal-to-noise ratio, using new contrast mechanism. The use of hyperpolarized noble gas, such as xenon-129 (5658) or helium-3 (5961), has been demonstrated. After the noble gas (xenon or helium) is inhaled, the airway and alveolar air space can be visualized with strong signal intensity, again demonstrated in both animals and humans.
As discussed above, one of the early limitations of clinical CT scanners was their ability to image only structure and not function, but with the increase in speed and improved spatial resolution along with the use of intravenous or intrabronchial contrast enhancement, physiologic information can be derived. Serial CT combined with nonradioactive xenon gas (atomic number, 54; K edge, 34.6 keV) has been used to measure cerebral blood flow by applying the theory proposed by Kety and Schmidt (62). Gur and coworkers (50, 63, 64) used CT with nonradioactive, radiodense xenon gas (Xe-CT) to measure regional pulmonary ventilation, taking advantage of the better spatial and temporal resolution of CT, compared with xenon scintigraphy. However, little attention was paid to careful lung volume control and the images suffered from significant misregistration problems when scans performed before and after xenon inhalation scans were subtracted from one another to visualize ventilation. Although powerful, Xe-CT techniques continue to face two primary shortcomings: (1) radiation dose is of growing public concern and there remains a need for methods that lower dose levels and (2) the need to track wash-in and/or wash-out of the gas at a fixed level of the thorax limits the current ability to image ventilation throughout the whole lung. The z-axis coverage for current MDCT scanners is at most 4 cm.
Two Xe-CT techniques for the measurement of regional pulmonary ventilation have been developed: single-breath and multibreath technique. Ball and coworkers (65) tried to measure regional ventilation by xenon scintigraphy by relating the count rate over the lung after a single breath to that after equilibration of the isotope throughout the lung. Tajik and coworkers (20) showed the potential of the single-breath method using Xe-CT, which was performed by having the subject take a single deep breath of a high nonradioactive xenon gas concentration mixture (80% xenon, 20% oxygen). Subtraction of identically gathered xenon-enhanced and unenhanced images can yield information regarding the relative distribution of local pulmonary ventilation. The enhancement due to xenon yields direct assessments of well-ventilated versus poorly ventilated lung regions. The single-breath technique has several advantages: it minimizes radiation dose and it allows for volumetric, spiral scanning. A limitation of the single-breath method lies in the fact that one must take a large breath of a high concentration of xenon. In some individuals this can cause the unwanted effect of temporarily eliminating the drive to breathe. Multibreath wash-in and/or wash-out methods offer information related to regional ventilation under dynamic conditions.
Xenon Wash-in and Wash-out Studies
X-ray CT assessments of local pulmonary ventilation have assumed monoexponential kinetics for xenon tracer gas filling or clearing the alveoli, and the model has been described by Kety (66).
Since xenon (Xe) gas was first introduced by Knipping and coworkers (67) as a contrast material to examine pulmonary function, it has been used extensively in scintigraphy (65, 68, 69), MRI (57), single-photon emission computed tomography (70), and X-ray analysis (57, 63, 65, 6871). Stable (nonradioactive) xenon gas has a K edge similar to that of iodine and should, therefore, be a potent X-ray attenuator and provide good contrast enhancement when used in conjunction with CT scanning (72). When imaged in a conventional CT scanner, lung attenuation varies linearly with the xenon concentration in the inhaled gas (73). As discussed by Tajik and coworkers (20), the slope of the relationship between attenuation and xenon concentration was 1.55 Hounsfield units (HU) per percentage increase in xenon concentration for the electron-beam CT scanner and 2.24 HU per percentage increase in xenon concentration for the spiral/axial CT scanner. Also, attenuation is dependent on the kilovolt setting, with lower kilovolt settings giving greater attenuation because of the physical properties of xenon gas (74). Xenon is moderately soluble in blood and tissue. The Oswald solubility coefficient is 0.14 at 37°C for blood and 0.13 for fat tissue (75). The solubility of xenon gas has been considered one source of errors in the application of multibreath wash-in and/or wash-out protocols.
Errors caused by xenon solubility in blood.
A number of investigators have used the Kety model in xenon scintigraphy (76, 77) and Xe-CT techniques (78), assuming xenon gas as the special case of gas with low solubility in blood. Previous studies using xenon scintigraphy reported that because of the solubility of xenon gas in blood and tissue, xenon is absorbed by the blood stream and distributed to the body parts during the wash-in phase.
Xenon wash-in and/or wash-out protocols.
In the application of Xe-CT, xenon wash-in, xenon wash-out, and combined xenon wash-in/wash-out have all been used to measure cerebral blood flow in both research and clinical studies. Polacin and coworkers (79) suggested a 3-min wash-in/5-min wash-out protocol by comparing it with an 8-min wash-in protocol and evaluating precision and accuracy of the cerebral blood flow estimates. For the measurement of regional pulmonary ventilation, three protocols (20, 71, 74) have been applied in both animal and human studies, based on the Kety model, which assumes that xenon is insoluble in blood and tissue and that the xenon wash-in rate is equal to the xenon wash-out rate. Simon and coworkers (80) used a Monte Carlo simulation technique to compare the estimates of confidence interval limit in three different protocols using the Xe-CT technique. The result from Monte Carlo simulation showed that the combined wash-in/wash-out protocol gives a narrower confidence interval for the time constant with the same nominal parameters and noise level than does either wash-in or wash-out alone. Tajik and coworkers (20) showed the repeatability of the regional tracer wash-out and wash-in measurement in terms of calculated time constants. Twenty regions of interest were randomly chosen on lung base. While repeatability of wash-out measurement was performed on the basis of data obtained with the EBCT scanner, wash-in measurement repeatability was performed with the spiral/axial CT scanner. Repeatability was compared with data obtained from only one animal in the supine position. The slope was 1.01 (R2 = 1.0) for EBCT and 1.00 (R2 = 0.96) for spiral/axial CT.
The use of wash-in and wash-out curves to derive a single estimate of regional ventilation has been brought into question. Chon and coworkers (81) have demonstrated that the two curves do not mirror each other. These investigators have hypothesized that, because of the density of xenon gas, as the gas enters the lung and meets the lighter resident gas, Rayleigh-Taylor instabilities cause churning effects that in turn cause nonequal distribution of the inhaled gas down bifurcating paths. This observation has raised questions concerning whether inhaled xenon gas will provide the same information as inhaled helium gas. Because xenon and helium are used increasingly to study regional ventilation via polarized gas methods associated with MRI (55, 82, 83) and xenon is of increasing interest as a marker of regional ventilation assessed via CT, it is important to understand the effect of gas density and viscosity on regional ventilation. Lin and Hoffman (84) have begun to use computational fluid dynamic modeling in conjunction with airway geometry assessed via CT imaging to gain insights into these newly emerging questions.
The density and viscosity of xenon gas are 5.44 g/L and 2.25 x 105 Pa/s at ambient temperature, respectively (85). These values are substantially higher than those of the other gases, which may impair respiratory function when used as an inhaled anesthetic. Previous studies showed the effect of physical gas properties on pulmonary mechanics and gas exchanges, revealing that respiratory mechanics are affected by the density and viscosity of a specific gas mixture. Zhang and coworkers (86) found an increase in airway resistance related to the inspiratory xenon concentration. Because of the particularly high density and viscosity of xenon gas, the effects of a xenonoxygen mixture on respiratory resistance are markedly sensitive to changes in the inspired xenon concentration. Rueckoldt and coworkers (87) reported a marked increase in peak air pressure and a 2.04-fold Reynolds number increase during inhalation of a 33% Xe67% O2 mixture, causing the zone of transition from turbulent to laminar gas flow to locate more peripherally with a consecutive increase in airway resistance.
We provide an example of the mapping of regional ventilation via Xe-CT in a prone sheep (Figure 3, left [p. 506]) along with a map of perfusion (Figure 3, middle) and V·/Q· (Figure 3, right) in the same sheep. Specific ventilation (per minute) is color coded. Chon and coworkers (81) have interrelated wash-in and wash-out time constants to gas density and lung structure, and Marcucci and coworkers (74) have demonstrated significant differences in ventilation patterns in the prone versus supine body posture.
VentilationPerfusion
Much of what has been known regarding the physiology of ventilation and perfusion matching has come from early scintigraphic imaging of radioactive xenon gas (11, 88, 89). A less direct method using a multiple inert gas elimination technique was introduced by Wagner and coworkers (14). With the introduction of Xe-CT to assess regional ventilation and CT methods for assessing regional perfusion, if one carefully images an individual at the same lung volume and body posture and adjusts small mismatches in anatomy occurring between the two scanning protocols via image-warping methods (90) it becomes possible to image regional ventilation and perfusion and to map voxel-based histograms of V·/Q· heterogeneity as discussed by Hoffman and coworkers (91). Kreck and coworkers (92) proposed a method for extracting regional V·/Q· from a single xenon wash-in study. This model takes into account the solubility of xenon gas in blood and tissue, leading to separate measurement of ventilation, perfusion, and a ventilationperfusion ratio. If the Hounsfield unit (X-ray attenuation coefficient) change for a given percentage of xenon gas is known, and the original regional xenon gas volume into which the inhaled xenon gas is being equilibrated, it is possible to estimate the regional density plateau that should be achieved during a wash-in maneuver. To the extent that the predicted regional density enhancement is not achieved, one can estimate regional perfusion with the assumption that the failure to achieve the predicted regional density enhancement is due to the gas being carried away by the blood. The limitation of this model is the assumption of a fixed "personal" (regional) dead space, the need for relatively high xenon radiodensity signal, and the low signal-to-noise characteristics of the perfusion signal (predicted minus actual plateau).
CONCLUSIONS
MDCT in conjunction with careful gating of scanning to either respiratory or cardiac signals is now providing the ability to link structure with function in unprecedented detail. It is expected that with this new information, we will begin to gain new insights into the etiology of disease processes, devise improved methods for disease intervention, better evaluate the efficacy of these interventions, and establish phenotypes that can aid in the search for genotypes associated with many common lung pathologies. With the growing interest in the use of mouse models to help understand the genetic basis of lung disease, micro-CT (93, 94) is emerging as a powerful new tool for imaging the lung at resolutions on the order of 9 to 18 µm, which allows for the visualization of anatomic structure at the level of the alveoli in the intact mouse. As shown in Figure 4, with careful gating of scanning and respiration, it is possible to image the mouse lung such that details of the lung microstructure are clearly visible, and accurate density measures allow for assessment of regional lung function directly related to gas delivery to the lung periphery and the status of the parenchyma at the gas exchange interface.
|
The authors thank all the members of the Iowa Comprehensive Lung Imaging Center (University of Iowa) who have made the work discussed here possible. The micro-CT image (Figure 4) was acquired in collaboration with Eman Namati.
FOOTNOTES
Supported by National Institutes of Health grants HL-064368 and HL-060158.
The color figures for this article are on p. 506.
Conflict of Interest Statement: E.A.H. is a co-owner of VIDA Diagnostics, which has marketing rights to image analysis software discussed in this article. D.C. does not have a financial relationship with a commercial entity that has an interest in the subject of this manuscript.
(Received in original form September 12, 2005; accepted in final form September 21, 2005)
REFERENCES
Related articles in Proceedings of the American Thoracic Society:
This article has been cited by other articles:
![]() |
J. P. Mizgerd and S. J. Skerrett Animal models of human pneumonia Am J Physiol Lung Cell Mol Physiol, March 1, 2008; 294(3): L387 - L398. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. B. Dolovich and D. P. Schuster Positron Emission Tomography and Computed Tomography versus Positron Emission Tomography Computed Tomography: Tools for Imaging the Lung Proceedings of the ATS, August 1, 2007; 4(4): 328 - 333. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Petersson, A. Sanchez-Crespo, S. A. Larsson, and M. Mure Physiological imaging of the lung: single-photon-emission computed tomography (SPECT) J Appl Physiol, January 1, 2007; 102(1): 468 - 476. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||